MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS BATTERY RECIPROCATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS BATTERY RECIPROCATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.715_LL
Device Problems Failure to Power Up (1476); Sticking (1597); Device Damaged by Another Device (2915)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and determined that the device did not function and had no power.It was further determined that the trigger was sticking.It was observed that the electric motor was damaged and not functional.It was further observed that there was an unknown liquid found on the internal components and the seals were worn and damaged.Therefore, the reported condition was confirmed.The assignable root cause associated with the trigger sticking was due to component wear and the unknown liquid and seal issue was due to improper cleaning/care, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported that the battery reciprocator device would not power up at all.According to the reporter, the ¿bad reciprocator¿ device damaged the battery device.During in-house engineering evaluation, it was observed that the trigger was sticking.The event was not reported to have occurred during surgery.It was not reported if there were any delays in a surgical procedure or if a spare device was available.The reporter stated that there was no patient incident reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY RECIPROCATOR II FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5939453
MDR Text Key54762696
Report Number8030965-2016-14796
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 07/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.715_LL
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2016
Date Manufacturer Received08/30/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/17/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
BATTERY DEVICE
-
-