Model Number TPO100B |
Device Problem
Device Alarm System (1012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional information become available, a supplemental record will be filed.
|
|
Event Description
|
Provider states the unit isn't alarming and there isn't any error on the screen.
|
|
Manufacturer Narrative
|
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the pcb had an open circuit, causing the alarms not to function and no error on the screen, which confirmed the original complaint issue.
|
|
Event Description
|
Provider states the unit isn't alarming and there isn't any error on the screen.
|
|
Search Alerts/Recalls
|