• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 840 VENTILATOR; CONTINUOUS VENTILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 840 VENTILATOR; CONTINUOUS VENTILATOR Back to Search Results
Model Number 840
Device Problems Device Emits Odor (1425); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The service engineer (se) verified the malfunction and replaced the analog interface (ai) printed circuit board (pcb).The unit passed all testing and operates within the manufacturing specifications.
 
Event Description
It was reported that an 840 ventilator had an alert message; breath delivery (bd) power on self-test (post) failure with burnt smell.The ventilator was not in use on a patient at the time the malfunction occurred.
 
Manufacturer Narrative
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI 
Manufacturer (Section G)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI  
Manufacturer Contact
ray maroofian
2101 faraday ave.
carlsbad, CA 92008
7606035334
MDR Report Key5942907
MDR Text Key54764425
Report Number8020893-2016-02354
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K970460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number840
Device Catalogue Number10049998
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received08/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-