Model Number 840 |
Device Problems
Device Emits Odor (1425); Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The service engineer (se) verified the malfunction and replaced the analog interface (ai) printed circuit board (pcb).The unit passed all testing and operates within the manufacturing specifications.
|
|
Event Description
|
It was reported that an 840 ventilator had an alert message; breath delivery (bd) power on self-test (post) failure with burnt smell.The ventilator was not in use on a patient at the time the malfunction occurred.
|
|
Manufacturer Narrative
|
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|