INVACARE FLORIDA OPERATIONS XPO2 PORTABLE CONCENTRATOR 9153648166; GENERATOR, OXYGEN, PORTABLE
|
Back to Search Results |
|
Model Number XPO100 |
Device Problems
No Audible Alarm (1019); Out-Of-Box Failure (2311)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional information become available, a supplemental record will be filed.
|
|
Event Description
|
Dealer is stating out of box failure, that there is no breath alarm will not activate, unit goes into an auto pulse mode without activation.
|
|
Manufacturer Narrative
|
Additional/updated information was added to reflect the concentrator being returned to the manufacturer for evaluation.The result of the evaluation was that the no breath alarm was not functioning, which confirmed the original complaint issue.The underlying cause could not be determined.However, once the unit was reset to factory specifications, the unit alarmed for no breath detected.
|
|
Event Description
|
Dealer is stating out of box failure, that there is no breath alarm will not activate, unit goes into an auto pulse mode without activation.
|
|