The patient was implanted with a left ventricular assist device (lvad).It was reported that during an outpatient clinic visit, the patient reported that there was a small cut in the silicon part of the driveline.The distributor's representative evaluated the driveline and noted some fluid underneath the silicon of the driveline.As it reportedly looked like soap, it was assumed to be residue after showering.A tape repair was performed.The following week, the patient reported a small new puncture just above the repaired point, where now blood was exiting from.All device parameters remained stable and the patient was reportedly feeling well.The cut in the driveline was covered with a sterile compress to let the blood drain out.On 08/22/2016, the distributor's presentative sealed off the puncture with silicon and closed it further with rescue tape.No further information was provided.
|