Model Number U6060 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Submit date: 09/15/2016.An investigation is currently under way, upon completion the results will be forwarded.
|
|
Event Description
|
It was reported to covidien on (b)(6) 2016 that a customer had an issue with a foot pump.The customer states low compressor opt.Upon triage on (b)(6) 2016 the service tech found the unit had a damaged power cord with exposed copper wire.
|
|
Manufacturer Narrative
|
Submit date: 10/10/2016.An investigation of av-impulse controller for the reported condition was performed for the reported condition of; exposed copper wire.The unit was triaged and the complaint was confirmed.The cause of the reported condition was due to rough handling.Power cords periodically require replacement due to age, usage and user damage.The service manual instructs the user to periodically inspect the power cords resistance to ensure its electrical safety.The unit was manufactured in 2008.A review of the device history record shows this device was released meeting all manufacturing specifications.
|
|