Model Number U6060
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 09/15/2016.An investigation is currently under way, upon completion the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2016 that a customer had an issue with a foot pump.The customer states low compressor opt.Upon triage on (b)(6) 2016 the service tech found the unit had a damaged power cord with exposed copper wire.
 
Manufacturer Narrative
Submit date: 10/10/2016.An investigation of av-impulse controller for the reported condition was performed for the reported condition of; exposed copper wire.The unit was triaged and the complaint was confirmed.The cause of the reported condition was due to rough handling.Power cords periodically require replacement due to age, usage and user damage.The service manual instructs the user to periodically inspect the power cords resistance to ensure its electrical safety.The unit was manufactured in 2008.A review of the device history record shows this device was released meeting all manufacturing specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
U6060 A-V IMPULSE CONTROLLER
Type of Device
FOOT PUMP
Manufacturer (Section D)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key5953778
MDR Text Key55259803
Report Number3008361498-2016-00089
Device Sequence Number1
Product Code JOW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberU6060
Device Catalogue NumberU6060
Device Lot NumberAVI80892
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received10/10/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-