Model Number IRC10LXO2 |
Device Problems
No Audible Alarm (1019); Loss of Power (1475)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional information become available, a supplemental record will be filed.
|
|
Event Description
|
The unit has been down for months.The unit beeps then shuts down with no audible alarm when turned on.
|
|
Manufacturer Narrative
|
Additional/updated information was added to reflect the device being returned to an independent repair center for evaluation.The result of the evaluation was that the 4-way valve was cracked, causing low o2/yellow light.However, there is no indication that there was a failure of the alarms.Therefore, this is no longer an fda reportable event.The following fields were updated accordingly:.
|
|
Event Description
|
The unit has been down for months.The unit beeps then shuts down with no audible alarm when turned on.Per the independent repair center, the reason for repair is low o2/yellow light.The key failure is the 4 way valve is cracked.
|
|