(b)(4).B.Braun (b)(4) became aware of a software error in software (sw) 8.2a of the dialog+ dialysis machine affecting the timer/stop watch function.This timer function only serves as a support to the user to remind of an action that has to be taken by the user.The failure occurs only in sw 8.2a and only in combination with option adimea.In certain conditions and on rare occasions, the warning triggered to remind the user is reset by the machine without user interaction.The user might not be aware that the reminder was triggered at all.The root cause of this failure is a single error in sw 8.2a.If the error occurs, the warning sounds and the yellow light flashes as intended, but the warning is reset by the machine without user interaction.If the timer/stop watch function is used to remind the user of the application or change of a medication during therapy, and the reminder fails, the patient might suffer from consequences dependent on the medication intended to be applied.Since the market introduction of the sw 8.2a in 2010 only one single case was reported recently in germany describing this situation.In this case the concentrate was not changed.The patient did not suffer from any long-term consequences.The failure has been corrected with software version 8.2b.All dialog+ dialysis machines with software 8.2a and option adimea must be updated to software 8.2b.Based on the investigation results of this complaint a voluntary recall had been initiated in the countries where the software version 8.2a of the dialog+ machine in combination with option adimea is marketed.Since the affected software version 8.2a is not marketed in the usa, it is not required to include the usa in this voluntary recall.
|