• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN TOMIOKA CORPORATION ZM-521PA; TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN TOMIOKA CORPORATION ZM-521PA; TRANSMITTER Back to Search Results
Model Number ZM-521PA
Device Problems Battery Problem (2885); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/19/2016
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the transmitter got very hot near the batteries and was not in use on a patient at the time.The device was sent to nihon kohden for evaluation, at which time the reported issue was duplicated.The customer was provided with an exchange device.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The customer reported that the transmitter got very hot near the batteries.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-521PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: susan shadley
tomioka city, japan 370-2 314
JA  370-2314
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: susan shadley
tomioka city, japan 370-2 314
JA   370-2314
Manufacturer Contact
susan shadley
1-31-4 nishiochiai, shinjuku-k
attn: susan shadley
tokyo, japan 161-8-560
JA   161-8560
2687133
MDR Report Key5959195
MDR Text Key55366803
Report Number8030229-2016-00456
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/18/2016,08/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZM-521PA
Device Catalogue NumberZM-231PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA09/18/2016
Distributor Facility Aware Date08/19/2016
Device Age18 MO
Event Location Hospital
Date Report to Manufacturer09/18/2016
Date Manufacturer Received09/18/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/05/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-