• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWIST DRILL,FLEX,CRVD, FOR 2.3 SA; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWIST DRILL,FLEX,CRVD, FOR 2.3 SA; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72203160
Device Problems Break (1069); Detachment Of Device Component (1104); Component Falling (1105)
Patient Problem Foreign Body In Patient (2687)
Event Date 05/24/2013
Event Type  malfunction  
Manufacturer Narrative
Visual inspection and functional testing could not be performed because the device in question was not returned for evaluation.Thus, the complaint could not be verified, nor could the root cause be determined with confidence.There are no indications that would suggest the device did not meet product specifications upon release into distribution.
 
Event Description
The tip broke off in the patient during a procedure utilizing a twist drill, flex, crvd, for 2.3 sa.It was removed by using an elevator, drilling around and pulled out using graspers.No patient injury or other complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWIST DRILL,FLEX,CRVD, FOR 2.3 SA
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
150 minuteman road
andover, MA 01810
5123585706
MDR Report Key5962897
MDR Text Key55449944
Report Number1219602-2016-00707
Device Sequence Number1
Product Code MBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/24/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Catalogue Number72203160
Device Lot Number50456025
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/24/2013
Date Device Manufactured05/02/2013
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-