MAUDE Adverse Event Report: CYBERONICS - HOUSTON PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS - HOUSTON PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Defective Device (2588); Programming Issue (3014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/01/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the nurse practitioner was experiencing issues getting her programming system to function.A company representative performed troubleshooting and isolated the issue to the serial cable.The representative found two faulty serial cables during the troubleshooting.The second cable was reported in mfg.Report #1644487-2016-02175.The physician was provided a new serial cable.The serial cable was received and underwent product analysis.During analysis, the returned serial cable was connected to a known good wand and tablet.The pa lab attempted to interrogate a known good generator with the serial cable however an "unable to open port" message presented.The serial cable's db9 hood was opened it was observed that a wire was no longer soldered on the pcb.The wire was soldered back into place.The lab then successfully interrogated a generator using the serial cable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING COMPUTER
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5973149
MDR Text Key55837690
Report Number1644487-2016-02176
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received08/30/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2015
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-