It was reported that the nurse practitioner was experiencing issues getting her programming system to function.A company representative performed troubleshooting and isolated the issue to the serial cable.The representative found two faulty serial cables during the troubleshooting.The second cable was reported in mfg.Report #1644487-2016-02175.The physician was provided a new serial cable.The serial cable was received and underwent product analysis.During analysis, the returned serial cable was connected to a known good wand and tablet.The pa lab attempted to interrogate a known good generator with the serial cable however an "unable to open port" message presented.The serial cable's db9 hood was opened it was observed that a wire was no longer soldered on the pcb.The wire was soldered back into place.The lab then successfully interrogated a generator using the serial cable.
|