• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37603
Device Problem Low impedance (2285)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A manufacturer representative (rep) reported that the patient is showing a low impedance on their left implantable neurostimulator (ins).Impedances for 0 and 1 were shown to be 33 ohms, with all other pairs within acceptable ranges.There were no patient symptoms alleged and follow up is being performed to obtain further information.Indication for implant is dystonia.
 
Event Description
Additional information was received from a manufacturer representative (rep).It was reported that the cause and the date of when the low impedances began occurring was unknown.It was noted, however, that they began occurring during the post-operative period.It was also unknown if the patient experience any symptoms associated with the issue.The rep stated that the actions/interventions taken to resolve the low impedances included programming around the low impedance pair.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key5973153
MDR Text Key55466066
Report Number3007566237-2016-03357
Device Sequence Number1
Product Code MRU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number37603
Device Catalogue Number37603
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-