Analysis of the instrument and instrument data indicate that the cause for the falsely depressed hb1c results is user error.The account failed to properly reconstitute the hb1c calibrators that are included with the hb1c kit.The calibrator instructions for use direct the customer to reconstitute the calibrators with 2 ml of water but the customer reconstituted with only 1 ml.A calibration was performed with the under-reconstituted vials and patient results were reported before the out of range qc values were detected.The issue was resolved with a proper preparation of calibrators, a recalibration, and confirmation with qc.Corrected patient results were issued.The device is performing within specifications.No further evaluation of the device is required.
|