Dr.(b)(6) of (b)(6) hospital used a m27/29 (sn (b)(4), expiry date: 2020-02-10) to perform a mvr two weeks ago.After finishing the sewing, he used the leaflet probe to test the movement of the leaflets.Dr.(b)(6) said there was one leaflet not moving properly.He used the rotator to turn the valve leaflets to another position, it did not improve the leaflet movement.Dr.(b)(6) finally decided to remove this m 27/29 from patient and replaced by another on-x m 27/29.The mvr was completed without affecting the quality of operation.Dr.(b)(6) complained about the quality of product.
|
A sample review was performed on the on-x mitral heart valve ¿ 27/29 (onxm-27/29, sn (b)(4)) on 10/21/2016 via gross visual examination and dimensional inspection.The valve met all dimension and functional inspections; it is likely that the suspect leaflet was prevented from moving by interference from the subvalvular retained mitral apparatus. the manufacturing records for the onxm-27/29, sn (b)(4), were reviewed and it was confirmed that all records were controlled, available for review, and met all specifications per the device master record.All lots passed functional testing and met release specifications. evidence gathered in the sample review concludes that the leaflet impingement was likely due to the physiological interference and is addressed in the ifu.The dfmea addresses excessive leakage under valve mechanics, listing one cause of failure as leaflet impingement.Additionally, the failure mode identified during the sample evaluation is a monitored event assessed in the bi-annual risk assessment for the on-x heart valve.This event does not identify additional hazards or modify the probability and severity of existing hazards.Evaluation of the returned valve indicated that the valve met specifications and concluded that the leaflet impingement was likely due to physiological interference such as subvalvular retained mitral apparatus.Additionally, the surgeon replaced the valve with another on-x valve of the same size, and the second valve did not present with leaflet motion issues.A review of manufacturing records indicates that all lots passed functional testing and met release specifications.
|