• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 701027652
Device Problems No Audible Alarm (1019); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 09/08/2016
Event Type  malfunction  
Manufacturer Narrative
Maquet medical systems, usa submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary (b)(4).A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
It was stated that during maintenance / service, after turned on the arterial pump "stop _ _" error message without audible alarm.No patient was involved.(b)(4).
 
Manufacturer Narrative
Field safety engineer has been sent for investigation.According to the service order no.(b)(4) he inspected and replaced the boards.Tested working ok.Performed function test and had pump running for several hours.Passed without any faults/errors.The replaced parts are not available for lab investigation.The failure could be confirmed.The most possible root cause could be determined as defective motor control board and control board.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HL-20 INTEGRATED PERFUSION SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key5977807
MDR Text Key55916881
Report Number8010762-2016-00596
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701027652
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/08/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-