Reportable based on device analysis completed on 30aug2016.It was reported that a guidewire kink occurred.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified left anterior descending (lad) artery.A rotawire¿ and wireclip¿ torquer was selected for use.Upon preparation, outside the patient's body, it was noticed that the rotawire was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status is good.However, device analysis revealed the wire was broken in the middle of the device.
|