Model Number 1050052
Device Problem No Flow (2991)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Reported device not marketed in the u.S., however, similar devices or devices that share components, (b)(6) process methods or any other technological characteristics with a medical device that is registered within the u.S.Aesculap inc.(importer) is submitting this report on behalf of (b)(4) (manufacturer).(b)(4).Evaluation on-going.Device not returned.
 
Event Description
(b)(6).Glue has no flow, tip was blocked.Facility reported multiple lot numbers with this issue.Three reports are filed: 2916714-2016-00727, 2916714-2016-00728, 2916714-2016-00729.
 
Manufacturer Narrative
Samples received: 90 closed and 3 open pouches.Analysis and results: there are no previous complaints of this code batch.(b)(4) units were manufactured and distributed in the market.There are no units in stock.Seven out of 30 samples evaluated were affected.There are polymerization nuclei along the cannula that caused the obstruction of the cannula.Ageing studies were performed with the purpose of provoking the complained defect and study the influence of environmental conditions in its apparition.-ampoules were stressed for 9 days at 40ºc and 30 % rh.After 4 days of exposure, the stress conditions caused the apparition of some polymerization nuclei along the cannula in the ampoules.However, the liquid glue could still pass through the cannula, that was not obstructed.After 9 days of exposure, the polymerization nuclei observed caused the obstruction of the cannula.The environmental conditions influence the apparition of the polymerization nuclei along the cannula.A temperature of 40ºc caused the formation of polymerization nuclei along the cannula and the obstruction of the cannula after 9 days of exposure.The instructions for use, histoacryl should be stored at ambient temperature below 22ºc.The ampoule containing the adhesive should only be removed from the aluminum pouch immediately prior to application.Final conclusion: taking into account that the samples received do not fulfill the oem specifications, it is concluded that the complaint is justified.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HISTOACRYL BLUE TISSUE ADHESIVE 0.5ML
Type of Device
TISSUE ADHESIVES
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, barcelona 08191
SP  08191
Manufacturer (Section G)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, barcelona 08191
SP   08191
Manufacturer Contact
nicole broyles
615 lambert pointe drive
hazelwood, MO 63042
3145515988
MDR Report Key5979742
MDR Text Key55705036
Report Number2916714-2016-00728
Device Sequence Number1
Product Code MPN
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
NARRATIVE
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2017
Device Model Number1050052
Device Catalogue Number1050052
Device Lot Number215402N2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date08/18/2016
Date Manufacturer Received10/20/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-