Model Number 48-40-00
Device Problem Improper Device Output (2953)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/30/2016
Event Type  No Answer Provided  
Manufacturer Narrative
Patient information was not provided.Sorin group (b)(4) manufactures the sorin s5 system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group received a report that the flow sensor of the sorin s5 system suddenly showed a drop in flow when the rpm was maxed during a procedure.There was no report of patient injury.A sorin group field service representative was dispatched to the facility to investigate.The customer reported to the service representative that no unusual resistance was found in the circuit.The service representative tested the unit and was unable to reproduce the reported issue.The flow sensor was replaced as a precaution and a functional verification was performed.No further issues were reported and the unit was returned to service.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the flow sensor of the sorin s5 system suddenly showed a drop in flow when the rpm was maxed during a procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN S5 SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w 65th.way
arvada, CO 80004
3034676461
MDR Report Key5985100
MDR Text Key55874470
Report Number9611109-2016-00675
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 08/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number48-40-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/30/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/14/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-