• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Partial Blockage (1065); Positioning Failure (1158)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/25/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during implant procedure, the left ventricular lead helix had difficulty extending.The physician attempted to flush the lead but was unsuccessful.The lead was not used and replaced successfully.
 
Manufacturer Narrative
Final analysis confirmed that it was difficult to insert the stylet into the lead.Dried saline was found at 22.7cm from the connector pin which prevented the stylet to be inserted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
15900 valley view court
sylmar, CA 91342
8184934022
MDR Report Key5992246
MDR Text Key56465227
Report Number2017865-2016-06332
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Model Number1458Q/86
Device Lot NumberA000019553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/25/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-