• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ON-X LIFE TECHNOLOGIES, INC ON-X AORTIC HEART VALVE WITH CONFORM-X SEWING RING AND EXTENDED HOLDER -25MM; MECHANICAL VALVE CONDUIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ON-X LIFE TECHNOLOGIES, INC ON-X AORTIC HEART VALVE WITH CONFORM-X SEWING RING AND EXTENDED HOLDER -25MM; MECHANICAL VALVE CONDUIT Back to Search Results
Model Number ONXACE-25
Device Problems Contamination (1120); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Endocarditis (1834)
Event Date 12/10/2015
Event Type  Injury  
Manufacturer Narrative
Note: the initial report for 1649833-2016-00016 submitted 09/16/2016 contained information for the wrong product and was updated in this follow-up report.In 2016, on-x technologies incorporated (on-x) became a wholly owned subsidiary of cryolife, inc.(cryolife).A retrospective review of the on-x complaint system was performed to identify any areas requiring additional action and to appropriately assimilate the on-x process into the cryolife quality management system.In an abundance of caution this mdr is being reported to satisfy regulatory reporting obligations.Operated valve endocarditis is a recognized risk of prosthetic valves that could lead to thrombosis, thrombotic embolism, bleeding events, or paravalvular leak [akins 2008].The objective performance criteria of iso 5840 for rigid prosthetic valves indicates an endocarditis rate of 1.2 %/pt-yr.In the proact study for avr , there were 9 reported cases out of 375 implants [puskas 2014).Endocarditis of prosthesis is an infrequent, but known risk of aortic valve replacement using prosthetic heart valves.Endocarditis is a known potential complication listed in the instructions for use (ifu).Furthermore, this complication is not unique to the on-x device and is a risk associated with all mechanical heart valves.All observed risks are mitigated in the labeling and ifu.No further action required.Without return of the valve, a definitive root cause cannot be identified.
 
Event Description
Implant recovery cards received indicate that patient implanted with onxace-25 (4824304, valve 1) on (b)(6) 2015 indicated for aortic valve regurgitation and bacterial endocarditis.Re-intervention and explant required on (b)(6) 2015 for bacterial endocarditis, replaced with onxace-23 (4945312, valve 2).Additional re-intervention/explant required on (b)(6) 2015 indicated for "4+ aortic insufficiency with a rocking aortic valve suggesting dehiscence" and replaced with onxace-23 (3926503).This report is relegated to valve 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-X AORTIC HEART VALVE WITH CONFORM-X SEWING RING AND EXTENDED HOLDER -25MM
Type of Device
MECHANICAL VALVE CONDUIT
Manufacturer (Section D)
ON-X LIFE TECHNOLOGIES, INC
1300 e. anderson ln.
bldg b
austin TX 78752
Manufacturer (Section G)
CRYOLIFE, INC.
1655 roberts blvd., nw
kennesaw GA 30144
Manufacturer Contact
rochelle maney
1655 roberts blvd., nw
kennesaw, GA 30144
7704193355
MDR Report Key5995201
MDR Text Key56333973
Report Number1649833-2016-00018
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier00851788001471
UDI-Public00851788001471
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date05/21/2020
Device Model NumberONXACE-25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/02/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age36 YR
-
-