• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Catalog Number 75646535
Device Problems Corroded (1131); Migration or Expulsion of Device (1395)
Patient Problem No Code Available (3191)
Event Date 09/08/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Product analysis: microscopic and sem examination of the set screws revealed severe pitting present on the set screw where they interface with the mas threads, as well as at the base of the set screw where it interfaces with the rod.This severe pitting could mechanically weaken the attachment points of the construct.Both the mas head and the set screw are made from astm f2229 stainless steel.Microscopic examination of the mas revealed pitting present at the base of the saddle/u-channel where it interfaces with the rod, as well as in the set screw interface threads.The pitted locations are around component interface points.The pitted areas seen under microscopic and sem examination appear to have a striated granular appearance with cubical voids, when viewed under very high magnification.These voids are indicative of chemical attack.The pitting seen at these construct interface points, when as sembled, provide crevices which can promote in vivo corrosion.Only the set screws were examined via sem (scanning electron microscope), due to sem chamber size constraints.Sem microscopy identified cuboidal pitting /voids seen on set screw rod interface surfaces.Complaint returned product corrosion location and surface morphology consistent with crevice corrosion.Microscopic examination of mas head identified crystallographic pitting in the rod saddle / mas head, consistent in location and surface morphology with crevice corrosion.Stainless steel corrosion resistance is obtained from a passive film which forms in an oxygen rich environment; the presence of large amounts of biological material may have generated oxygen deficient regions, resulting anodic cell development, which would result in the loss of the passive film and eventual corrosion of the material.Additionally, micro-motion of the components may have also contributed by mechanically removing the passive film formed on the stainless steel.No other material signs of wear, cracking, fracture or breakage were identified.The location pitted surface morphology and microstructures are consistent with anticipated wear due to crevice corrosion.
 
Event Description
It was reported that the patient underwent a surgery due to scoliosis at t2-l2.Post-op, the patient had bent over and felt a "pop" in her back.X-rays noticed that the pedicle screw was pulling out.Revision surgery was done to remove, re-instrument and add additional levels.The implant did not break but corrosion was seen in the implant.The patient did not achieve solid fusion.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Type of Device
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
greg anglin
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key5995939
MDR Text Key56332911
Report Number1030489-2016-02743
Device Sequence Number1
Product Code KWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number75646535
Device Lot NumberH13C3207
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ROD
Patient Outcome(s) Required Intervention;
Patient Age19 YR
-
-