Device Problems
Device Emits Odor (1425); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).The service engineer (se) inspected the device and was not able to confirm the burning smell however, a review of the diagnostic logs confirmed the alleged serial number mismatch.The se replaced the universal serial bus (usb).The se performed extended self-testing on the device and all tests passed.
|
|
Event Description
|
It was reported that, while in use on a patient, a 980 ventilator had a burning smell, generated serial number mismatch and went into an inoperative state.The patient was removed from the ventilator and manually ventilated via an ambu bag before being placed on an alternate ventilator.The patient was not harmed or injured as a result of the reported event.
|
|
Manufacturer Narrative
|
An investigation was performed and the technician found that the reported problem could not be duplicated.No failure was detected.If information is provided in the future, a supplemental report will be issued.
|
|