COVIDIEN 980 VENTILATOR; CONTINUOUS VENTILATOR Back to Search Results
Model Number 980
Device Problems Device Emits Odor (1425); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
(b)(4).The service engineer (se) inspected the device and was not able to confirm the burning smell however, a review of the diagnostic logs confirmed the alleged serial number mismatch.The se replaced the universal serial bus (usb).The se performed extended self-testing on the device and all tests passed.
 
Event Description
It was reported that, while in use on a patient, a 980 ventilator had a burning smell, generated serial number mismatch and went into an inoperative state.The patient was removed from the ventilator and manually ventilated via an ambu bag before being placed on an alternate ventilator.The patient was not harmed or injured as a result of the reported event.
 
Manufacturer Narrative
An investigation was performed and the technician found that the reported problem could not be duplicated.No failure was detected.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI 
Manufacturer (Section G)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI  
Manufacturer Contact
ray maroofian
2101 faraday ave
carlsbad, CA 92008
7606035334
MDR Report Key5996661
MDR Text Key56376219
Report Number8020893-2016-02604
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131252
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number980
Device Catalogue Number980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received09/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-