Model Number BE-PLS 2050
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/08/2016
Event Type  malfunction  
Manufacturer Narrative
The device has been requested but not yet received.A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
During priming, a black particle was found on the outlet side of the oxygenator.For that reason, this product couldn¿t be used.(b)(4).
 
Manufacturer Narrative
The product was investigated in the laboratory of the manufacturer.A grey foreign matter was detected on the blood inlet side beneath the de-airing membrane.In order to identify the particle the oxygenator was sawed and the foreign matter was separated.The particle was grey and rubber-like.The dimensions are: length 3,64 mm, width 1,61 mm.The source and the identity of the particle is unknown.Based on this the failure "foreign matter on upper venous side" could be confirmed.Trend search: a sap trend search was performed for p/n 70102.7818, failure code 0719 foreign matter and no additional complaint was found.Due to this information no systemic issue could be determined.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/BIOLINE COATING
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6000919
MDR Text Key57115485
Report Number8010762-2016-00609
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
PMA/PMN Number
K080592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBE-PLS 2050
Device Catalogue Number701027818
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-