Catalog Number 237081
Device Problem Premature Activation (1484)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/02/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation at this time.The manufacturer will continue to monitor and trend related events.
 
Event Description
The clips fell from the applier.The patient's condition was reported as fine.
 
Manufacturer Narrative
(b)(4).The dhr for the instrument in question was reviewed and found completely without any irregularities.This instrument was produced at the teleflex.Medical (b)(4) facility in april of 2012 as part of a (b)(4) pc.Lot.Since the instrument was not returned for evaluation, we are unable to validate this complaint or determine root cause at this time.All instruments are thoroughly inspected and function tested at time of manufacture.No corrective action taken at this time.
 
Event Description
The clips fell from the applier.The patient's condition was reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HZ APPLIER MED 8" CVD
Type of Device
APPLIER,SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key6002917
MDR Text Key56582525
Report Number3011137372-2016-00311
Device Sequence Number1
Product Code GDO
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number237081
Device Lot Number06F1149241
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/03/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-