• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC.; CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC.; CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR) Back to Search Results
Model Number 5487
Device Problem Material Integrity Problem (2978)
Patient Problem Cardiac Arrest (1762)
Event Date 08/03/2016
Event Type  Injury  
Manufacturer Narrative
Product event summary: manufacturer's analysis was unable to confirm the customer comment that the cable would not allow pacing and capture.The cable passed all visual incoming inspection with no anomalies found.The cable also passed continuity testing with no intermittent or shorted connections found.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the temporary pacing cable was connected to a patient's lead and pacing and capture did not occur.The patient experienced long periods of asystole as a result.The cable was replaced and returned for analysis.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6009480
MDR Text Key56782903
Report Number2182208-2016-02503
Device Sequence Number1
Product Code DSA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960446
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5487
Device Catalogue Number5487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/03/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age24 YR
-
-