Catalog Number 295025-001
Device Problems Material Rupture (1546); Material Protrusion/Extrusion (2979)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 09/14/2016
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although the freedom onboard battery has a protruding connector piece, it did not prevent the freedom driver from performing its life-sustaining functions because patients are provided with several onboard batteries, and the freedom driver has a redundant power source of external wall power.The freedom onboard battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer reported that one of the patient's freedom onboard batteries has a protuding connector piece.
 
Manufacturer Narrative
The freedom onboard battery was returned to syncardia for evaluation.Physical inspection of the onboard battery revealed housing separation, a broken and exposed j1 connector and an impact dent adjacent to the damaged area, confirming the customer-reported issue.The likely root cause for the damage is impact shock, as evidenced by the impact dent.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
 
Event Description
The customer reported that one of the patient's freedom onboard batteries has a protuding connector piece.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM ONBOARD BATTERY
Type of Device
BATTERY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6011766
MDR Text Key57285842
Report Number3003761017-2016-00323
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295025-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/09/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age49 YR
-
-