This is a continuation of the reported high impedance associated with mfr.Report #1644487-2016-01582.The patient was undergoing generator replacement surgery on (b)(6) 2016, and high impedance was identified when the new generator was attached to the existing lead.The lead was then replaced, and the high impedance resolved.However, programming data from the surgery was reviewed, and there was evidence of incomplete pin insertion causing fluctuating impedance from normal to high impedance.Also, programming data from the patient's previous generator showed that there was no high impedance prior to the generator being replaced, and analysis of the lead showed evidence of possible incomplete pin insertion into the generator, as reported in mfr.Report #1644487-2016-01582.Therefore, the cause of the high impedance was incomplete pin insertion.No further relevant information has been received to date.
|