• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET ORTHOPEDICS COMPREHENSIVE REVERSE SHOULDER GLENOSHPERE BASEPLATE; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS COMPREHENSIVE REVERSE SHOULDER GLENOSHPERE BASEPLATE; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problem Packaging Problem (3007)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/15/2016
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly.Examination of returned device found the product to be non-conforming to print.Review of the device confirmed the reported condition, as a hair was noted inside the shrink wrap of the device.The most likely root cause was operator error, as the packaging was not inspected according to process specification.Further investigation was initiated to address the reported issue.It was deemed no further actions necessary.
 
Event Description
It was reported that during a shoulder procedure, a hair was discovered in the shrink wrap of the glenoid baseplate.Another device was used to complete the procedure.
 
Manufacturer Narrative
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPREHENSIVE REVERSE SHOULDER GLENOSHPERE BASEPLATE
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key6015539
MDR Text Key56988985
Report Number0001825034-2016-04049
Device Sequence Number1
Product Code PAO
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
PK080642
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number115330
Device Lot Number055510
Other Device ID NumberSEE NARRATIVE IN H10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/25/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-