MAUDE Adverse Event Report: THERAKOS, INC THERAKOS CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS, INC THERAKOS CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Lot Number E131-KIT
Device Problems Fluid/Blood Leak (1250); Increase in Pressure (1491); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/15/2016
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.A batch record review of kit lot e131 was conducted.There were no nonconformances associated with this lot.The lot met release requirements.The uvadex lot number was not provided as it was not administered.However, a review of all uvadex lots manufactured since january 2013 was performed.No trends or nonconformances related to the complaint were noted.Trends were reviewed for complaint categories, pressure dome membrane leak, alarm #16: collect pressure, and alarm #17: return pressure.No trends were detected for these complaint categories.The assessment is based on information available at the time of the investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.(b)(4).Device not returned to manufacturer.
 
Event Description
The customer called to report a pressure dome membrane leak at the system pressure sensor.The customer stated that the leak occurred after several alarm #16: collect pressure and alarm #17: return pressure alarms.The customer reported that the treatment was aborted with no blood/products returned to the patient.The customer stated that the patient was in stable condition and had left the hospital after the procedure.The customer will follow-up with the patient and will call back in case there are any issues.The kit was not returned for investigation as it had already been discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERAKOS CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX
Manufacturer (Section D)
THERAKOS, INC
hampton NJ
Manufacturer (Section G)
THERAKOS, INC.
10 north high street
suite 300
west chester PA 19380
Manufacturer Contact
megan vernak
53 frontage road
hampton, NJ 08827
MDR Report Key6019592
MDR Text Key57619463
Report Number2523595-2016-00227
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Nurse
Type of Report Initial
Report Date 09/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date06/01/2018
Device Lot NumberE131-KIT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/15/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-