It was reported the patient had gone to the er on (b)(6) 2016 complaining of dyspnea, coughing, and a tingling sensation in her face and upper extremities.The physician tested the vns and found high impedance.The vns was programmed off and the patient was referred for surgery.The patient underwent surgery to correct the high impedance observed, on (b)(6) 2016.It was noted that the lead pin was not fully inserted into the generator.The lead pin was reinserted; the impedance was checked both outside and inside the generator pocket.Both sets of diagnostics were within normal limits and there was no need for a replacement of either the lead of the generator.
|