Submit date: 11/01/2016.A review of the information in the complaint file indicates this investigation was performed by a medtronic technical center for the reported condition of; a power issue.During evaluation a damaged power cord with exposed copper wire was found.Therefore, this report will be based on information provided by the technical center.The service finding for the power cord with exposed copper wire was confirmed.The cause of the reported condition was due to misuse.The damaged power cord was scrapped and replaced to correct the reported issue.The unit passed all initial testing after failure analysis repair instructions were completed.The 700 series scd was manufactured in 2011.A review of the device history record shows this device was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.Based on the investigation, no corrective action is needed.
|