• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CPI - DEL CARIBE ENDOTAK RELIANCE; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CPI - DEL CARIBE ENDOTAK RELIANCE; IMPLANTABLE LEAD Back to Search Results
Model Number 0184
Device Problems Fracture (1260); Over-Sensing (1438); Pocket Stimulation (1463); Inappropriate/Inadequate Shock/Stimulation (1574); Low impedance (2285); Ambient Noise Problem (2877)
Patient Problem Muscle Stimulation (1412)
Event Date 09/06/2016
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that the implantable cardioverter defibrillator (icd) and right ventricular (rv) lead exhibited low out of range pacing impedance measurements of less than 200 ohms along with muscle stimulation and oversensing of noise that led to inappropriate shocks.It was noted that the noise was able to be reproduced with isometrics.The physician suspected a lead fracture.A revision procedure was performed where the rv lead was surgically abandoned and replaced.The icd remained in service with the new rv lead.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key6023998
MDR Text Key57247908
Report Number2124215-2016-15473
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/14/2011
Device Model Number0184
Other Device ID NumberENDOTAK RELIANCE G
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received09/06/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/14/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0184; 4469; E110
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age82 YR
Patient Weight82
-
-