• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE 1288HD CAMERA CONTROL UNIT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE 1288HD CAMERA CONTROL UNIT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 1288010000
Device Problem Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/15/2016
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.(b)(4).
 
Event Description
It was reported that smoke was coming from the device prior to surgery.
 
Manufacturer Narrative
Alleged failure: smoking, burning overheating.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be defective component on the digital board.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.(b)(4).
 
Event Description
It was reported that smoke was coming from the device prior to surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1288HD CAMERA CONTROL UNIT
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
kimberly lynch
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key6025997
MDR Text Key57321513
Report Number0002936485-2016-00964
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132785
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number1288010000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received09/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/02/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-