Catalog Number 1288010000 |
Device Problem
Smoking (1585)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/15/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.(b)(4).
|
|
Event Description
|
It was reported that smoke was coming from the device prior to surgery.
|
|
Manufacturer Narrative
|
Alleged failure: smoking, burning overheating.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be defective component on the digital board.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.(b)(4).
|
|
Event Description
|
It was reported that smoke was coming from the device prior to surgery.
|
|