TELEFLEX MEDICAL HUDSON THERMAGARD NEBULIZER HEATER; BREATHING SYSTEM HEATER Back to Search Results
Catalog Number 5705
Device Problems Insufficient Heating (1287); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/11/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device sample was not returned for evaluation at the time of this report.
 
Event Description
The customer alleges that the device stopped heating during use.The patient's condition is reported as fine.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and it was observed that the heater was received with the ring adaptor.It was found that the power switch would light up, but the unit would not generate heat.Based on the investigation performed, the reported complaint was confirmed.The unit is approximately 720 days old.The heater was found to be abused by immersion in some type of liquid, and all the inner components were damaged.It was determined that the unit was beyond economical repair.
 
Event Description
The customer alleges that the device stopped heating during use.The patient's condition is reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON THERMAGARD NEBULIZER HEATER
Type of Device
BREATHING SYSTEM HEATER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key6033180
MDR Text Key57622586
Report Number3011137372-2016-00322
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number5705
Device Lot NumberSN 19372
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received11/10/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-