Please note "usage of device" should be reuse.The returned external pacemaker was thoroughly analyzed.It was principally capable to pace and sense.However, it was found during the analysis that the external damages to the operating elements did not allow for a proper operation.Damage and soiling was found.The housing is cracked several times and fractured, and all setting wheels are hard to turn because of the dirt.The visual and mechanical analysis, as well as the function check, did not show any indications of a material defect or manufacturing error.The device was damaged by external impacts.
|