Device manufacture date: 2015.(b)(6).The field service specialist (fss) visited customer site to inspect the unit.Fss checked and found the rear panel controller (rpc) printed circuit board assembly (pcb) had burned out and the power distribution cable was faulty.Fss replaced the rear panel controller and power distribution cable assembly.Fss then checked for all parameter as per field service check list, which all values were found to be within the limits as per amo specifications.Through follow ups, fse confirmed that there was no other visible evidence of fire/shock/explosion and no visible smoke was observed by him.That, there was no charring/melting of the power distribution cable; the cable was faulty due to connection at the inner point, and the burn had only happened on the rear panel controller pcb.All pertinent information available to abbott medical optics has been submitted.
|
The device manufacture date of 2015 was listed in the initial mdr.Included in the device evaluation information it was indicated that the device manufacturing date was 07/2015.The labeling, trending and risk documentation reviews for this equipment were performed.The trend review shows that there is not a recognizable adverse trend.The risks and mitigations associated with the received complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.A labeling review was conducted; the operator manual for the system was reviewed and found to include adequate instructions for use, warnings and operational errors.The review of the device history record (dhr) for this system was also performed which showed that there were no issues or non-conformities and that the system and its components met all specifications prior to being released.All pertinent information available to abbott medical optics has been submitted.
|