Model Number H749236310030 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/22/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that the brake defeat did not function.During use of a 1.50mm rotalink¿ plus, the brake was not holding the rotawire completely in place.The wire was held in place manually to complete the procedure.The patient was not affected.There were no patient complications and the patient is in stable condition.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: returned product consisted of the rotablator rotalink plus device.The advancer unit and burr units were received attached together in a single unit.The advancer knob was received loosened in a backward position.There was blood on and in the advancer unit.The brake button and advancer were microscopically and visually inspected and no damage or irregularities were identified.The brake button was received in the as-manufactured position.The brake button was tested and it was able to be pressed and released with no issues.The guidewire used in the procedure was not returned for product analysis, so functional testing was completed with.009¿ rotawire.The rotawire was able to be inserted in the rotablator plus device and the brake worked properly with no issues.The advancer was opened and the brake was examined.There was no damage or irregularities to the brake or the inside the advancer at the brake location.Inspection of the device presented no damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
|
Event Description
|
It was reported that the brake defeat did not function.During use of a 1.50mm rotalink¿ plus, the brake was not holding the rotawire completely in place.The wire was held in place manually to complete the procedure.The patient was not affected.There were no patient complications and the patient is in stable condition.
|
|
Search Alerts/Recalls
|