UNOMEDICAL LTD. UNOMETER¿ SAFETI¿ PLUS - WITH LUER-LOCK & LUER; DEVICE, URINE FLOW RATE MEASURING, NON-ELECTRICAL, DISPOSABLE
|
Back to Search Results |
|
Model Number 158100140190 |
Device Problem
Disconnection (1171)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/04/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Expiration date (12/2020).Manufacturing date (01/2016).Based on the available information, this event is deemed a reportable malfunction.No patient harm was reported.Additional information has been requested but not received to date.When additional information becomes available, a follow-up report will be submitted.Reported to the fda on: october 20, 2016.(b)(4).
|
|
Event Description
|
Complaint reporting that "the nurse was going to empty the measuring chamber and as the blue lever was opened, the catheter disconnected from the measuring chamber, splashing the urine on the nurse's clothing." the use of the device was suspended, no additional treatment was required.No further details have been provided.
|
|
Manufacturer Narrative
|
Device history records were reviewed.Sterilization was performed in accordance with parameters.The batch was released according to requirements.A batch record review indicates no non-conformances are related to the complaint issue.The batch record review resulted in a discrepancy which was investigated in another complaint and is closed.On the basis of information received the investigation concludes that the true root cause for the issue "the inlet tube is detached from the measurement chamber" cannot be identified.No additional investigation is needed.A photograph has been received for this complaint, which was evaluated.This issue will be monitored through the post market product monitoring review process.No additional patient/event details have been provided to date.Should additional information become available, a follow-up report will be submitted.Reported to the fda on november 03, 2016.(b)(4).
|
|
Search Alerts/Recalls
|
|
|