• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNOMEDICAL LTD. UNOMETER¿ SAFETI¿ PLUS - WITH LUER-LOCK & LUER; DEVICE, URINE FLOW RATE MEASURING, NON-ELECTRICAL, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNOMEDICAL LTD. UNOMETER¿ SAFETI¿ PLUS - WITH LUER-LOCK & LUER; DEVICE, URINE FLOW RATE MEASURING, NON-ELECTRICAL, DISPOSABLE Back to Search Results
Model Number 158100140190
Device Problem Disconnection (1171)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/04/2016
Event Type  malfunction  
Manufacturer Narrative
Expiration date (12/2020).Manufacturing date (01/2016).Based on the available information, this event is deemed a reportable malfunction.No patient harm was reported.Additional information has been requested but not received to date.When additional information becomes available, a follow-up report will be submitted.Reported to the fda on: october 20, 2016.(b)(4).
 
Event Description
Complaint reporting that "the nurse was going to empty the measuring chamber and as the blue lever was opened, the catheter disconnected from the measuring chamber, splashing the urine on the nurse's clothing." the use of the device was suspended, no additional treatment was required.No further details have been provided.
 
Manufacturer Narrative
Device history records were reviewed.Sterilization was performed in accordance with parameters.The batch was released according to requirements.A batch record review indicates no non-conformances are related to the complaint issue.The batch record review resulted in a discrepancy which was investigated in another complaint and is closed.On the basis of information received the investigation concludes that the true root cause for the issue "the inlet tube is detached from the measurement chamber" cannot be identified.No additional investigation is needed.A photograph has been received for this complaint, which was evaluated.This issue will be monitored through the post market product monitoring review process.No additional patient/event details have been provided to date.Should additional information become available, a follow-up report will be submitted.Reported to the fda on november 03, 2016.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNOMETER¿ SAFETI¿ PLUS - WITH LUER-LOCK & LUER
Type of Device
DEVICE, URINE FLOW RATE MEASURING, NON-ELECTRICAL, DISPOSABLE
Manufacturer (Section D)
UNOMEDICAL LTD.
zavodskaya street
50 fanipol dzerzhinsk district
minsk region, minskaya voblasts 22275 0
BO  222750
Manufacturer Contact
jeanette johnson
7900 triad center drive
suite 400
greensboro, NC 27409
3365424681
MDR Report Key6045630
MDR Text Key58026228
Report Number3007966929-2016-00095
Device Sequence Number1
Product Code FFG
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number158100140190
Device Lot Number220378
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/25/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age56 YR
Patient Weight54
-
-