TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
|
Back to Search Results |
|
Model Number 6382 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Invalid Sensing (2293)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/26/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Per additional information from the distributor, the flow sensor was showing a flow rate when the line was clamped off, no specific values were reported.There was no visual damage to the flow sensor, cable or connector noted.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the flow sensor was reporting inconsistent readings.As a result, the customer used another sensor to complete the case successfully.There was no delay, no blood loss, nor adverse consequences reported.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.Per additional information from the distributor, no part will be returned to the manufacturer since the device started working fine after the customer cleaned it.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|