Model Number 6382
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Invalid Sensing (2293)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/26/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Per additional information from the distributor, the flow sensor was showing a flow rate when the line was clamped off, no specific values were reported.There was no visual damage to the flow sensor, cable or connector noted.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the flow sensor was reporting inconsistent readings.As a result, the customer used another sensor to complete the case successfully.There was no delay, no blood loss, nor adverse consequences reported.
 
Manufacturer Narrative
The reported complaint was confirmed.Per additional information from the distributor, no part will be returned to the manufacturer since the device started working fine after the customer cleaned it.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL SYSTEM
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
eileen dorsey
6200 jackson road
ann arbor, MI 48103
7347416074
MDR Report Key6045881
MDR Text Key58415083
Report Number1828100-2016-00693
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6382
Device Catalogue Number6382
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/31/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/10/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-