DEPUY SYNTHES POWER TOOLS 5.0CM SHORT ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT Back to Search Results
Catalog Number SHORT
Device Problems Break (1069); Migration or Expulsion of Device (1395); Component Missing (2306); Device Operates Differently Than Expected (2913); Naturally Worn (2988)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: (b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported by (b)(6) that the attachment device had an undetermined malfunction.During the pre-repair diagnostics assessment, it was determined that the device had damaged component to the bearings.It was further determined that the crani strap was worn out and the extension sleeve was slightly damaged.It was further observed that the heavily woven pins had moved (migrated) and the symbol was missing.It was further determined that the device failed for visual assessment and for vibration.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0CM SHORT ATTACHMENT
Type of Device
MOTOR, DRILL, ELECTRIC - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key6053474
MDR Text Key58662572
Report Number1045834-2016-13051
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K011444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 07/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSHORT
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2016
Date Manufacturer Received10/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/31/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-