During a generator replacement, the new generator's impedance was 604 ohms, which the surgeon considered low.Preoperatively, on the explanted generator, a dcdc-0 was observed upon diagnostics.The surgeon examined the lead and found that it was frayed.The lead was explanted and replaced.A company representative present at the surgery reported that blood was found in the lead.Replacing the lead resolved the issue of low impedance.Review of the lead device history records confirmed that the lead passed all quality specifications prior to distribution with no non-conformances.Product analysis was performed on the lead.The majority or the lead body, including the lead pins, electrode array, and tie downs, was not returned to the manufacturer.On the portion returned, both the inner and outer lead tubing was found to be abraded with the inner quadfilar coils exposed.Dried bodily fluids were found in the inner and outer lead tubing.There were multiple places along the outer tubing where abrasions were found, indicating possible wear.These findings support the low impedance and fluid leaks reported by the surgeon.The patient's physician reported that patient manipulation or trauma to the lead could have possibly contributed to the lead abrasions.No additional relevant information has been received to date.
|