Catalog Number BLACKMAX-N
Device Problems Air Leak (1008); Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 10/06/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This is report 2 of 2 of the same event.It was reported that during a discectomy surgical procedure, it was observed that the craniotome device broke into two and the motor device sounded like it had an air leak.According to the report, the two devices were being used together when the event occurred.There was an unspecified delay to the surgical procedure.It was unknown if there were spare devices available for use.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device passed all manufacturing specifications.Therefore, the reported condition was not duplicated and confirmed.An assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLACKMAX-NEURO
Type of Device
MOTOR, DRILL, PNEUMATIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key6054773
MDR Text Key58607684
Report Number1045834-2016-13052
Device Sequence Number1
Product Code HBB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K831756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberBLACKMAX-N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received11/09/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
CRANIOTOME DEVICE
-
-