The cause for the false high total hcg result is unknown.A siemens field service engineer (fse) was sent to the customer site for system inspection on 10/04/2016.Siemens is awaiting further information from the customer site.The ifu states in the interpretation of results section: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." the ifu states in the limitations section: "if an aberrant or abnormal result, as defined by the laboratory protocol, occurs, laboratory personnel should first make certain that the system is performing and is operated and maintained in accordance with the product labeling.The user should then follow the laboratory protocol for advising the clinician of a result that appears to have deviated from the norms established by the laboratory.Test results alone are not diagnoses of medical conditions.For example, low titer elevations of hcg can occur in normal nonpregnant subjects.A physician's diagnosis involves evaluation of the test result in conjunction with, and in the context of, the patient's medical history, physical examination, and other test results, sometimes in consultation with other medical experts.".
|
Siemens filed the initial mdr 1219913-2016-00188 on october 26, 2016.On 12/01/2016 additional information: the siemens field service engineer (fse) report was provided.The fse found, during the verification of the vacuum circuit, a presence of a plug in the small y connector between the water trap and the waste reserve (fluidic schematics).The fse removed the jam/plug and cleaned the connector.The cause for the false high total hcg result appears to be due to the jam/plug in the connector.The instrument is performing within specifications.No further evaluation of the device is required.Advia centaur xpt instrument information: (b)(4), 510k #: k041133, product code: jje, catalog #: 10711433.
|