• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. EXTRA CURVED BROACH HANDLE; HIP INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. EXTRA CURVED BROACH HANDLE; HIP INSTRUMENT/TRIAL Back to Search Results
Catalog Number 259807550
Device Problem Mechanical Problem (1384)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
This complaint is under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The broach handle has become loose.
 
Manufacturer Narrative
This complaint remains under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.
 
Manufacturer Narrative
A functional test with a mating broach confirmed the complaint; the handle no longer locks.The root cause is attributed to misuse and wear; damage to the cam component.It is important to ensure that the alignment features on both the handle and broach are mated correctly, as proper locking will extend the life of all broach handles.If these orientation features are not aligned properly, the cam will lock incorrectly onto the broach.After this type of deformation occurs, the handle will no longer lock as tightly onto the broach as when first distributed.Based on the investigation, the need for corrective action is not indicated.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRA CURVED BROACH HANDLE
Type of Device
HIP INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6057477
MDR Text Key58409571
Report Number1818910-2016-30225
Device Sequence Number1
Product Code HTQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 12/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number259807550
Device Lot NumberRA129A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/12/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-