• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. SIG HP PATELLA & INSERTION CSE; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. SIG HP PATELLA & INSERTION CSE; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 950502808
Device Problem Delamination (2904)
Patient Problem No Information (3190)
Event Date 08/29/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Examination of the submitted instrument cases confirmed the brackets are delaminating.A search of the complaint database found additional reports of bracket delamination for sigma hp instrument cases against the complaint sample product codes.Previous investigation did not conclusively determine the root cause, although it is the supplier manufacturing process related.Details of this issue have been documented through pra / hhe (dva- (b)(4)) and capa (b)(4).Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Event Description
The plastic coating peeling off the sigma hp instrument trays from repeated wash/sterilization processes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG HP PATELLA & INSERTION CSE
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6058057
MDR Text Key58415057
Report Number1818910-2016-30227
Device Sequence Number1
Product Code FSM
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 08/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number950502808
Device Lot NumberDV3BF4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-