• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CARBOFLO EPTFE VASCULAR GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CARBOFLO EPTFE VASCULAR GRAFT Back to Search Results
Catalog Number F8006TWSC
Device Problem Torn Material (3024)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/28/2016
Event Type  malfunction  
Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device is not available for return.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that after implantation of the vascular graft an alleged split was identified approximately 3/4 of the way down from the suture site; therefore, the torn section of the graft was cut and removed.It was further reported that the remainder of the graft was re sutured, and the procedure was completed.There was no reported patient injury.
 
Manufacturer Narrative
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only event reported to date for this lot number and failure mode.Visual inspection & functional/performance evaluation: the device was not returned; therefore, no visual inspection or functional testing of the device was performed.Medical records review: medical records were not provided for review.Image/photo review: images/photos were not provided for review.Conclusion: the investigation was inconclusive for torn material, as the sample was not returned for evaluation.The definitive root cause could not be determined based upon available information.It was unknown whether patient and/or procedural issues contributed to the event.Labeling review: the current instruction for use (ifu) states: eptfe grafts do not stretch (are non-elastic) in the longitudinal direction.The correct graft length for each procedure must be determined by considering the patient's body weight, posture, and the range of motions across the anatomical area of graft implantation.Failure to cut the graft to an appropriate length may result in anastomotic or graft disruption, leading to excessive bleeding, and loss of limb or limb function, and/or death.Aggressive and/or excessive graft manipulation when tunneling, or placement within a too tight or too small tunnel, may lead to separation of the spiral beading and/or graft breakage.When suturing, avoid excessive tension on the suture line, inappropriate suture spacing and bites, and gaps between the graft and host vessel.Failure to follow correct suturing techniques may result in suture hole elongation, suture pull-out, anastomotic bleeding and/or disruption.Additional mfr narrative: date received by mfr; (conclusion); the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that after implantation of the vascular graft an alleged split was identified approximately ¾ of the way down from the suture site; therefore, the torn section of the graft was cut and removed.It was further reported that the remainder of the graft was re sutured, and the procedure was completed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARBOFLO EPTFE VASCULAR GRAFT
Type of Device
EPTFE VASCULAR GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer Contact
judith ludwig
1625 w 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key6059392
MDR Text Key58456044
Report Number2020394-2016-01006
Device Sequence Number1
Product Code DYF
UDI-Device Identifier00801741021626
UDI-Public(01)00801741021626(17)201113(10)VTZK0436
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K004012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/13/2020
Device Catalogue NumberF8006TWSC
Device Lot NumberVTZK0436
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/01/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-