• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE OEC MEDICAL SYSTEMS (SLC) 9900; FLUOROSCOPIC X-RAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE OEC MEDICAL SYSTEMS (SLC) 9900; FLUOROSCOPIC X-RAY Back to Search Results
Model Number 9900
Device Problem Loss of Data (2903)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/04/2016
Event Type  malfunction  
Manufacturer Narrative
A ge service representative performed an onsite investigation.The hard disk drive was evaluated and reformatted.The system software and calibrations were also reloaded.The system was tested and found to be working as intended and returned to service.
 
Event Description
It was reported that the system failed to properly save patient image data.No patient serious injury or death was reported related to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9900
Type of Device
FLUOROSCOPIC X-RAY
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers drive
salt lake city UT 84116
Manufacturer (Section G)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers drive
salt lake city UT 84116
Manufacturer Contact
384 wright brothers drive
salt lake city, UT 84116
MDR Report Key6064987
MDR Text Key58715343
Report Number1720753-2016-02949
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 10/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9900
Device Lot NumberES-3223
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-