CYBERONICS - HOUSTON PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problems Energy Output To Patient Tissue Incorrect (1209); No Device Output (1435)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/06/2016
Event Type  malfunction  
Manufacturer Narrative
Suspect udi: (b)(4).
 
Event Description
Information was received that a burst watchdog warning was seen by a healthcare professional.The warning indicated that the patient's device was programmed off because the burst watchdog warning had been triggered.The patient's magnet mode output current was equal to the programmed autostim current allowed a burst watchdog timeout to occur.This caused the patient's device to disable itself to 0ma output current but the exact time this occurred is not known.The physician has since altered the settings to avoid another scenario.No further relevant information has been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6067362
MDR Text Key59033003
Report Number1644487-2016-02476
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 10/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/09/2017
Device Model Number106
Device Lot Number4407
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received10/06/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/13/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age11 YR
-
-