• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 3116
Device Problems Unintended Collision (1429); Pocket Stimulation (1463); Inappropriate/Inadequate Shock/Stimulation (1574); Impedance Problem (2950)
Patient Problems Undesired Nerve Stimulation (1980); Pain (1994); Electric Shock (2554)
Event Type  Injury  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A heath care professional (hcp) via a manufacturer representative reported a consumer was getting stimulation in the pocket site.It was unknown when this started.The hcp noted impedances were "jumping around' and an endoscopy showed a bulge indicating a potential lead migration.The hcp was planning on replacing the system.Additional information received from the representative reported the cause of the potential migration of the lead was unknown and the issue had not been resolved.Information received from the consumer reported experiencing painful shocking along where the leads were, which started mid (b)(6) 2016.The shocking sensation was getting worse.She had followed up with her hcp, who performed multiple procedures with the consumer (x-ray, ct scan) and it was determined she needed to have her whole system replaced.They had been working with the representative in the area and had not heard back from him yet; they were waiting to schedule the replacement surgery.The consumer noted she had fallen on her implant about three weeks.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key6069599
MDR Text Key58847565
Report Number3004209178-2016-22964
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2013
Device Model Number3116
Device Catalogue Number3116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/04/2016
Date Device Manufactured05/31/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age43 YR
-
-