Model Number H74939407SO |
Device Problems
Bent (1059); Entrapment of Device (1212); Flaked (1246); Difficult to Remove (1528); Material Distortion (2977)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/04/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not received for analysis; therefore no physical or visual analysis of the product can be performed.The manufacturing batch record review confirmed that the device met all material, assembly and inspection specifications prior to shipment.Based on the information provided by the supporting documentation, procedural and/or clinical factors appear to have impacted on the event as reported.The guidewire is expected to be returned for analysis.If additional information is received or the product is returned for analysis a follow-up medwatch report will be submitted.No product returned.
|
|
Event Description
|
This mdr is for the first of the two guidewires mentioned below.A safari small and a lotus valve 27 mm were used.When the lotus valve system was outside the introducer, with the guidewire inserted into, the lotus valve system remained blocked without the possibility to move backward and forward.The guidewire has been cut close to the nosecone.A new safari small and a new lotus valve system 27 mm have been used.The valve implantation was successfully completed.During catheter removal and when the catheter was outside the introducer, the catheter remained blocked.Gently pulling, the catheter was removed completely and successfully.At visual inspection, a damage of the safari small guidewire was noticed.
|
|
Manufacturer Narrative
|
This is a follow-up to a previous medwatch report.The manufacturing batch record review confirmed that the device met all material, assembly and inspection specifications prior to shipment.The partial device was returned for analysis.As received, the specimen consisted of the proximal 33.75cm segment of one each safari2 275cm sml crv 1pk; returned loose and double-bagged within "zip-lock" style poly biohazard pouches.The complaint of damage was confirmed.The specimen presented several segments of damaged catheter sheath lodged onto the shaft and several bends and multiple offsets scattered over the length of the specimen.The specimen segment also presented extensive ptfe coating damage scattered over the bulk of the segment.The proximal coil to core wire joint is secure, despite the coil damage and cut damage to the external surface of the weld joint.It was not possible to assign a definitive root cause for the event as reported.Based on the evidence presented by the sample and the information provided by the supporting documentation, device interface and clinical factors appeared to have impacted on the event as reported.
|
|
Search Alerts/Recalls
|
|